Non-Contact Forehead Thermometer

TH30F

  • Non-contact for Hygiene and Comfort
  • One Second Measurement
  • Temperature Awareness Signal
  • Surface Mode & Room Temperature
  • Last Reading - Temperature Comparison
  • 25 Memories
  • °C/°F Switchable
  • TH30F
    Non-contact for Hygiene and Comfort
  • TH30F
    1 Second Measurement
  • TH30F
    Temperature Awareness Signal
  • TH30F
    Room Temperature
  • TH30F
    Last Reading
  • TH30F
    Surface Mode

Introduction

    • Temperature measureme range:  
      Foread mode:  34~42.2°C (93.2~108°F)  
      Surface mode:  -22~80°C (-7.6~176°F)
    • Accuracy:
      Forehead mode: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range.
      Surface mode: ±0.3°C (0.5°F) within  22~42.2°C (71.6~108°F), others ±4% or ±2°C (4°F) whichever is greater.
    • Operating temperature range:  10~40°C (50~104°F), 15%~85% RH 
    • Storage temperature:  -20~50°C (-4~122°F), RH ≤85%
    • Transportation temperature:  Shall be less than 70°C, RH ≤95%
    • Response time (90%):  
      Forehead mode: 1 sec.   
      Surface mode: 0.6 sec.
    • Display mode:  Oral / Surface
    • Scale selection:  °C / °F
    • Display type:  Large LCD
    • Memory:  25 sets
    • Temperature Awareness Signal:  Beep Sounds
    • Battery type:  CR2032 * 1 pcs
    • Dimension:  80.1 * 46.4 * 37.7 mm
    • Weight(with battery):  46.3 g
    Note: Subject to upgrade based on purchasing agreement.

    ► Feature
     
    TH48F_scan-distance

    Scan Distance ≤ 3CM

    last reading-th30f

    Last Reading

    TH48F_1sec

    One Second
    Measurement

    Easy Carry

    Easy Carry



      ► Certificates

    Certificates

     

Technology Support

    • 1.Why do I get different temperature for continuous measurement? Sometimes in normal, sometimes in lower, sometimes in higher temperature.+

      The user and thermometer need to adapt to ambient temperature. The rapid temperature change may cause the inaccurate measurement.

      User
      Before the measurement, please stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.

       Thermometer
      We recommend user to place thermometer indoor for 15 minutes to adapt to temperature and get the correct measurement.
    • 2.The measurement in air conditioning room is different from that in non- air conditioning room?+

      When ear thermometer is under rapid temperature change, we recommend user to place thermometer for 15 minutes to adapt to temperature and get the right temperature.
    • 3.What is the normal body temperature?+

      A normal range for body temperature is 35.8°C~38.0°C (96.4°F~100.4°F) . Each person has different temperature, please contact qualified doctor for details.

      Here are the normal body temperatures for reference:
      • Ear Temperature:35.7°C ~ 37.5°C
      • Rectal temperature:35.7°C ~ 37.5°C
      • Oral temperature:35.5°C ~ 37.5°C
      • Armpit temperature:34.7°C ~ 37.3°C


      Body temperature variations affected by age:
      • 0 ~ 2 years:36.4°C ~ 38.0°C
      • 3 ~ 10 years:36.1°C ~ 37.8°C
      • 11 ~ 65 years:35.9°C ~ 37.6°C
      • > 65 years:35.8°C ~ 37.5°C

      Source:
      1. Chamberlain, J.M., Terndrup, T.E., New Light on Ear Thermometer Readings, Contemporary Pediatrics, March 1994.
      2. Chamberlain, J.M., et al., Determination of Normal Ear Temperature with an Infrared Emission Detection Thermometer, Annals of Emergency Medicine, January 1995.
    • 4.Does forehead thermometer need to maintain regularly?+

      clean_IFT

      Forehead thermometer is a health care product that needs maintenance, you could use soft cloth to wipe the  appearance. Never use benzene for cleaning and never submerge thermometer into water or any liquids. Probe lens is the most delicate part of thermometer, please use cotton swab with alcohol (70%) to wipe it.

      Follow the below instruction and video to clean the probe lens.
      ►Probe Lens Cleaning Video

      forehead_cleaning_e(320x180)
       
      1. Before using the cotton swap with 70% alcohol to wipe the lens, slightly flip the cotton swab to cut down the spare alcohol.
      2. If there is no 70% alcohol, use alcohol prep pad to soak the cotton swab.

        Important:  Do not use sharp pointed object to wipe the lens.
Cookies
We use cookies to improve your user experience and for web traffic statistics purposes. By continuing to use this website, we’ll assume you agree to our use of cookies. Our Privacy & Cookies Policy contains more information on such use and explains how to disable cookies.
Accept