Multi-Function Thermometer
TH50Z
- Ear mode+Forehead mode
- One Second Measurement
- Temperature Awareniss Signal
- Surface Mode|Room Temperature
- Stop Watch|Real Time|Date
- Detailed 9 memories with time|date
- °C/°F Switchable
- Special CAP and Shape design to carry conveniently, and become larger size to hold well when measuring
Introduction
-
- Temperature measureme range:
Ear / Forehead mode: 34~42.2°C (93.2~108°F)
Surface mode: -22~80°C (-7.6~176°F) - Accuracy:
Ear / Forehead mode: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range.
Surface mode: ±0.3°C (0.5°F) within 22~42.2°C (71.6~108°F), others ±4% or ±2°C (4°F) whichever is greater. - Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature: -20~50°C (-4~122°F), RH ≤85%
- Transportation temperature: Shall be less than 70°C, RH ≤95%
- Response time (90%):
Ear mode: 1 sec.
Forehead mode: 1 sec.
Surface mode: 0.6 sec. - Display mode: Oral / Surface
- Scale selection: °C / °F
- Display type: Large LCD
- Memory: 9 sets
- Temperature Awareness Signal: Beep Sounds
- Battery type: CR2032 * 1 pcs
Note: Subject to upgrade based on purchasing agreement.► FeatureScan Distance ≤ 1CM
Zone Indicator
3-Color LED lightOne Second Measurement
Small Size Large Size
► Certificates
- Temperature measureme range:
Technology Support
-
-
-
-
-
1.Why is my measurement too low or high?+
You may check by following items:
- Ear Canal:
Earwax might block the infrared measure accuracy. - Probe Cover:
The most popular situation is the probe stuck with earwax that blocked infrared transmission and lower the measurement. If you bought a probe cover model, we recommend user to change probe cover each time to avoid cross contamination. - Note:
If you bought a probe cover model, the probe cover may break and affect the accuracy after using for several times or receive external hit. We recommend user to check probe cover's integrity before every measurement. If the probe cover is broken, please change a new one. - Lens:
If you bought a probe cover model, the most common situation is user measure without installing probe cover. The probe lens is stuck with earwax and lower the measurement. If the lens is dirty, please follow the manual's instruction to clean lens with cotton swab. - Note:
User hit the thermometer carelessly might cause the probe lens damage. Please return full package to retailer for service if there is damage.
- Ear Canal:
-
2.Why do I get different temperature for continuous measurement? Sometimes in normal, sometimes in lower, sometimes in higher temperature.+
The user and thermometer need to adapt to ambient temperature. The rapid temperature change may cause the inaccurate measurement.
User
Before the measurement, please stay in a stable environment for 5 minutes and avoid exercise, bath for 30mins.
Thermometer
We recommend user to place thermometer indoor for 15 minutes to adapt to temperature and get the correct measurement. -
3.What is the standard for fever?+
Different site in the body and different thermometers may get different temperatures.
Recommended fever standards stated below:
Site Fever Standard Measuring Time Oral 37.5 °C 1 min. half - 2 min. Forehead 37.5 °C Several seconds Ear 38.0 °C Several seconds Axillary 37.0 °C 3 min. Rectal 38.0 °C 1 min. - 1 min. half
Source ► Centers for Disease Control , R.O.C.(Taiwan) -
4.Does the forehead thermometer I bought only measure forehead temperature?+
RADIANT INNOVATION forehead thermometer is designed for forehead temperature measurement. Besides, you may also scan other subject's exterior temperature. For example, you can measure baby's milk or bath water in "SCAN" mode.
-