Non-Contact Forehead Thermometer
THD2FE
- Non-Contact for Hygiene and Comfort
- One Second Measurement
- Temperature Awareniss Signal
- Mute Function
- Surface Mode
- LED Backlight
- 25 Memories
- °C/°F Switchable
Introduction
-
- Temperature measureme range:
Foread mode: 34~42.2°C (93.2~108°F)
Surface mode: -22~80°C (-7.6~176°F) - Accuracy:
Forehead mode: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range.
Surface mode: ±0.3°C (0.5°F) within 22~42.2°C (71.6~108°F), others ±4% or ±2°C (4°F) whichever is greater. - Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature: -20~50°C (-4~122°F), RH ≤85%
- Transportation temperature: Shall be less than 70°C, RH ≤95%
- Response time (90%):
Forehead mode: 1 sec.
Surface mode: 0.6 sec. - Display mode: Oral / Surface
- Scale selection: °C / °F
- Display type: Large LCD
- Memory: 25 sets
- Temperature Awareness Signal: Beep Sounds
- Battery type: AAA * 2 pcs
Note: Subject to upgrade based on purchasing agreement.► FeatureScan Distance 1~4cm
Mute mode
One Second Measurement
Blue LED Backlight
► Certificates
► NoteThis product is certified as a medical device in the European Union under the Medical Device Regulation 2017/745 by SGS CE1639, exclusively for the indication(s) of non-contact forehead temperature measurement. Other non-medical uses ascribed to this device are not within the scope of CE certification, and users should be aware product performance and/or safety has not been evaluated by SGS for those purposes.
- Temperature measureme range:
Technology Support
-
-
-
-
-
1.Why is my measurement too low or high?+
You may check by following items:
- Ear Canal:
Earwax might block the infrared measure accuracy. - Probe Cover:
The most popular situation is the probe stuck with earwax that blocked infrared transmission and lower the measurement. If you bought a probe cover model, we recommend user to change probe cover each time to avoid cross contamination. - Note:
If you bought a probe cover model, the probe cover may break and affect the accuracy after using for several times or receive external hit. We recommend user to check probe cover's integrity before every measurement. If the probe cover is broken, please change a new one. - Lens:
If you bought a probe cover model, the most common situation is user measure without installing probe cover. The probe lens is stuck with earwax and lower the measurement. If the lens is dirty, please follow the manual's instruction to clean lens with cotton swab. - Note:
User hit the thermometer carelessly might cause the probe lens damage. Please return full package to retailer for service if there is damage.
- Ear Canal:
-
2.Does forehead mode need to touch body?+
Please measure close to "the temples" about 0.5 cm distance.
Hold the thermometer 1 to 3 cm from "the central forehead".
Although the thermometer does not touch body, please also follow the instruction for clean and storage to keep accuracy. -
3.Does summer and winter ambient temperature affect the measurement?+
RADIANT INNOVATION ear thermometer compares ambient temperature and makes compensation. The measurement between 10°C~40°C (50°F~104°F) is the same, but the human body changes a little in different season and different environment. For example, the body temperature becomes hotter after exercise and hot shower. -
4.Does the forehead thermometer I bought only measure forehead temperature?+
RADIANT INNOVATION forehead thermometer is designed for forehead temperature measurement. Besides, you may also scan other subject's exterior temperature. For example, you can measure baby's milk or bath water in "SCAN" mode.
-