Introduction
-
- Temperature measureme range: 34~42.2°C (93.2~108°F)
- Accuracy: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range
- Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature: -20~50°C (-4~122°F), RH ≤85%
- Transportation temperature: Shall be less than 70°C, RH ≤95%
- Response time (90%): 1 sec.
- Display mode: Oral
- Scale selection: °C / °F
- Display type: Large LCD
- Memory: 25 sets
- Disposable probe cover: Mold ring W/PE film cover
- Temperature Awareness Signal: Beep Sounds
- Battery type: CR2032 * 1 pcs
Note: Subject to upgrade based on purchasing agreement.► FeatureProbe Cover
LED Backlight
Probe Cover Detector
One Second Measurement
► Certificates
-
Do not reuse : Always use a clean, undamaged and new probe cover. Be sure ear canal is clean.
For RADIANT® Infrared Ear Thermometer "THXX6 series".Insert the probe into a new probe cover.Push the probe cover to the bottom of the probe until the “Click” sound. This means the probe cover has been installed successfully.
Technology Support
-
-
-
-
1.My temperature measured in the right ear is different from that measured in left ear?+
The ear canal of both ears is not totally the same. Some people can get same temperature but most people could get 0.5 ~ 1.0°C different temperatures. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature. -
2.The measurement in air conditioning room is different from that in non- air conditioning room?+
When ear thermometer is under rapid temperature change, we recommend user to place thermometer for 15 minutes to adapt to temperature and get the right temperature. -
3.What is the standard for fever?+
Different site in the body and different thermometers may get different temperatures.
Recommended fever standards stated below:
Site Fever Standard Measuring Time Oral 37.5 °C 1 min. half - 2 min. Forehead 37.5 °C Several seconds Ear 38.0 °C Several seconds Axillary 37.0 °C 3 min. Rectal 38.0 °C 1 min. - 1 min. half
Source ► Centers for Disease Control , R.O.C.(Taiwan) -
4.Does summer and winter ambient temperature affect the measurement?+
RADIANT INNOVATION ear thermometer compares ambient temperature and makes compensation. The measurement between 10°C~40°C (50°F~104°F) is the same, but the human body changes a little in different season and different environment. For example, the body temperature becomes hotter after exercise and hot shower.
-