Introduction
-
- Temperature measureme range: 34~42.2°C (93.2~108°F)
- Accuracy: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range
- Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature: -20~50°C (-4~122°F), RH ≤85%
- Transportation temperature: Shall be less than 70°C, RH ≤95%
- Response time (90%): 1 sec.
- Display mode: Oral
- Scale selection: °C / °F
- Display type: Large LCD
- Memory: 9 sets
- Disposable probe cover: Mold ring W/PE film cover
- Temperature Awareness Signal: Beep Sounds
- Backlight: Option
- Battery type: CR2032 * 1 pcs
Note: Subject to upgrade based on purchasing agreement.► FeatureMiniature Probe Engineered
(option)Probe Cover Detector
Probe Cover Ejector
(option)One Second
Measurement
► Certificates
-
For RADIANT® Infrared Ear Thermometer "THXX9 series".
Do not reuse : Always use a clean, undamaged and new probe cover. Be sure ear canal is clean.
(1) Attach the probe cover to the connection ring.
(2) Align the center of probe cover with the center of probe.
(3) Push the connection ring until the click sound. This means the probe cover has been firmed tightly.► Probe Cover Loader (option)Fill with probe covers
Pull the loader once to load a new probe cover
The loader will automatically bounce back. A new probe cover will be on the fixer.
Install a new probe cover
Technology Support
-
-
-
-
1.My temperature measured in the right ear is different from that measured in left ear?+
The ear canal of both ears is not totally the same. Some people can get same temperature but most people could get 0.5 ~ 1.0°C different temperatures. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature. -
2.The measurement in air conditioning room is different from that in non- air conditioning room?+
When ear thermometer is under rapid temperature change, we recommend user to place thermometer for 15 minutes to adapt to temperature and get the right temperature. -
3.What is the standard for fever?+
Different site in the body and different thermometers may get different temperatures.
Recommended fever standards stated below:
Site Fever Standard Measuring Time Oral 37.5 °C 1 min. half - 2 min. Forehead 37.5 °C Several seconds Ear 38.0 °C Several seconds Axillary 37.0 °C 3 min. Rectal 38.0 °C 1 min. - 1 min. half
Source ► Centers for Disease Control , R.O.C.(Taiwan) -
4.Does ear thermometer need to maintain regularly?+
Ear thermometer is a health care product that needs maintenance, you could use soft cloth to wipe the appearance. Never use benzene for cleaning and never submerge thermometer into water or any liquids. Probe lens is the most delicate part of thermometer, please use cotton swab with alcohol (70%) to wipe it.
►Probe Lens Cleaning Video
- Before using the cotton swap with 70% alcohol to wipe the lens, slightly flip the cotton swab to cut down the spare alcohol.
- If there is no 70% alcohol, use alcohol prep pad to soak the cotton swab.
Important: Do not use sharp pointed object to wipe the lens.
-