Infrared Ear Thermometer

TH819SJE

  • Probe Cover for Hygiene
  • One Second Measurement
  • Probe Cover Detector
  • Probe Cover Ejector
  • Worldwide Smallest Probe
  • Temperature Awareness Signal
  • LED Backlight
  • 9 memories
  • °C/°F Switchable
  • TH819SJE
    Probe Cover
  • TH819SJE
    1 Second Measurement
  • TH819SJE
    Temperature Awareness Signal
  • TH819SJE
    Wordwide Smallest Probe
  • TH819SJE
    Probe Cover Ejector
  • TH819SJE
    LED Backlight

Introduction

    • Temperature measureme range:  34~42.2°C (93.2~108°F)
    • Accuracy:  ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range
    • Operating temperature range:  10~40°C (50~104°F), 15%~85% RH 
    • Storage temperature:  -20~50°C (-4~122°F), RH ≤85%
    • Transportation temperature:  Shall be less than 70°C, RH ≤95%
    • Response time (90%):  1 sec.
    • Display mode:  Oral
    • Scale selection:  °C / °F
    • Display type:  Large LCD
    • Memory:  9 sets
    • Disposable probe cover:  Mold ring W/PE film cover
    • Temperature Awareness Signal:  Beep Sounds
    • Battery type:  CR2032 * 1 pcs
    Note: Subject to upgrade based on purchasing agreement.

     ► Feature
     
    th816e_probe-cover3

    Worldwide Smallest Probe

    th816e_led-backlight

    Probe Cover Ejector

    detector-th8

    Probe Cover Detector

    1sec

    One Second Measurement

     

     
     ► Certificates


     
  •  
    軟耳套安裝_new
    For RADIANT® Infrared Ear Thermometer "THXX9 series". 

    Donotreuse Do not reuse : Always use a clean, undamaged and new probe cover. Be sure ear canal is clean. 

    (1) Attach the probe cover to the connection ring.
    (2) Align the center of probe cover with the center of probe.
    (3) Push the connection ring until the click sound. This means the probe cover has been firmed tightly.
     
     

     
    ► Probe Cover Loader (option)
     
    step1

    Fill with probe covers

    step2

    Pull the loader once to load a new probe cover

    step3

    The loader will automatically bounce back. A new probe cover will be on the fixer.

    step4

    Install a new probe cover

Technology Support

    • 1.How to install the probe cover?+

      Please follow the instructions below.  Probe Cover Installation ►
      Install probe cover
      ( PC840 / PCL-A40 / PC7200 )

      (1)  Place a new probe cover on the connection ring. (See Figure 1)
      NOTE:Make sure to place the “Adhesive Side” of probe cover “Upward.”
       
      (2)  Align the probe with the center of probe cover. Insert the probe into the probe cover on the connection ring. (See Figure 2)
      (3)  Push the connection ring until the “Click” sound. This means the probe cover has been installed successfully.

      NOTE:The probe cover detector is optional. If the probe cover did not connect firmly, the “ SP_ear_clip_image002 ” icon will flash on the LCD screen.
      Please check the setting of probe cover again.


      軟耳套安裝_new_faq


      Warning:Keep the probe covers and connection ring away from children.
    • 2.How to get a precise measurement?+

      Check the five steps to get the best measurement:
       
      • STEP 1 :  Is the ambient temperature stable?
      • STEP 2 :  If you bought a probe cover model, is the probe cover a new one?
      • STEP 3 :  Is ear canal clear without earwax?
      • STEP 4 :  Is the ear canal straight during the measurement?
      • STEP 5 :  Did you take out thermometer after “Beep” sound?
    • 3.Is the ear temperature more precise than oral temperature?+

      The ear canal of both ears is not totally the same. Some people can get same temperature but most people get 0.5 ~ 1.0°C different temperatures. RADIANT INNOVATION consider the 0.4°C inaccuracy is due to the sensor did not completely receive the ear's infrared. Basically, when the self diagnosis function is normal, the different temperature is affected by the measuring probe's location. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature.
    • 4.Does ear thermometer radiate infrared?+

      The ear thermometer receives the ear drum's infrared and switch to body temperature; therefore, it does not radiate infrared.
Cookies
We use cookies to improve your user experience and for web traffic statistics purposes. By continuing to use this website, we’ll assume you agree to our use of cookies. Our Privacy & Cookies Policy contains more information on such use and explains how to disable cookies.
Accept
Search